I. Product Introduction
ES9020 intelligent lightning protection component tester, also known as lightning protection component tester, SPD field tester, used for arrester, lightning protector, surge protector, varistor, cermet discharge tube, direct space detonator and other performance test instrument. The instrument adopts the offline function of the charging function to overcome the shortcomings of inconvenient on-site operation of traditional products to be plugged in. The voltage output 2000V is more widely used, uses high-voltage short-circuit protection, automatic discharge function, and the meter is reliable, safe and durable. At the same time with insulation resistance test function, the use of luxury large-screen display, data storage, data review, compliance judgment, automatic shutdown, USB data upload and other functions. The machine is beautiful and upscale, has a wide range, high resolution, convenient operation, easy to carry, accurate, reliable, stable performance, strong anti-interference ability.
ES9020 also has a shockproof, dustproof, moisture-proof structure to adapt to harsh working environments. It is a commonly used and indispensable instrument for meteorological lightning protection, electricity, post and telecommunications, communications, electromechanical installation and maintenance, and industrial enterprises that use electricity as industrial power or energy. . It is suitable for measuring performance parameters of various arresters and surge protectors.
ES9020 smart lightning protection component tester consists of medium and large-scale integrated circuits. The output power of the meter is large, the initial operating voltage measurement range is 0-2000V, the leakage current measurement range is 0.1~199.9uA, the insulation resistance is 0-1000MΩ, the repeatability is good, and one-key measurement is convenient and quick.
II. Technical Specification
Function |
Varistor (MOV) initial voltage leakage current test, discharge tube (GDT) discharge voltage test, insulation resistance test |
Base condition |
23℃±5℃,below 75%rh |
Voltage measurement(V) |
0~2000V Accuracy±2%rdg±3dgt |
Leakage current measurement(uA) |
0.1uA~199.9uA Accuracy±3%rdg±5dgt |
DC voltage resolution |
1V |
Leakage current resolution |
0.1uA |
Insulation resistance |
0~1000MΩ |
Test settling time |
≤3s |
Discharge rate of discharge tube |
(100±8) V/s |
Power |
7.2V Large capacity rechargeable battery |
Backlight |
Yes, suitable for use in dim places |
USB data transmission |
Have |
Display mode |
5-inch color TFT LCD |
LCD display size |
108mm×65mm |
Instrument size |
240mm (L)×188mm (W)×85mm (H) |
Test line |
Test line red 1m, black 1m |
Data storage |
500groups |
Data review |
Data review function:“MR”symbol display |
Overflow display |
Over-range overflow function: "OL" symbol display |
Judging function |
Measurement exceeds the setting range |
Power consumption |
Standby:150mA Max |
Measure:880mA Max |
Weight |
1230g(including) |
Battery voltage |
Insufficient battery voltage, low battery symbol “ ”display |
Automatic shut-down |
Instrument will automatic shut-down after 15 minutes no operation |
Insulation resistance |
≥50MΩ (1000V) |
Pressure resistance |
AC3kV/50Hz 1min |
Working temperature and humidity |
-10℃~+50℃<85%RH |
Working temperature and humidity |
-15℃~+55℃<90%RH |
Suitable for Safety Regulations |
IEC61010-1、IEC1010-2-31、IEC61557-1,5、IEC60529(IP54)、pollution grate 2、CAT Ⅲ 300V |
上一篇:ES3035E high voltage insulation resistance tester (10KV, 1TΩ